Laboratory Glassware Washer also need to cooperate with cleaning agents and pay attention to regular maintenance

A well-designed Laboratory Glassware washer has a powerful circulation pump and well-designed nozzles.The cleaning solution can be evenly and continuously sprayed onto surface of the utensils to remove residue.It is true that many residue can be washed away by heat,water’s own ability to dissolve residue and spray pressure.
However,due to the high surface tension of water,the cleaning ability of pure water is limited for some tiny particles and organic substances that are difficult to dissolve in water.Due to the circulation and spraying of the automatic laboratory cleaning machine,ordinary cleaning agents contain surfactants,which will form a large amount of foam.On the one hand,these foams will overflow,on the other hand,it may also cause damage to the circulation pump,so the automatic laboratory cleaning machine can only use non-foaming cleaning agents.
The special cleaning agent contains not only alkali or acid,but also various active substances such as chelating agen and complexing agent.Through the synergistic effect of theses active substances,the residue can be better dissolved and dispersed.In addition,the cleaning solution must not only have the cleaning ability to remove residues,but also must not damage the surface and pipeline of the equipment.When laboratory glassware washer manufactures recommend cleaning agents, they have undergone careful testing and evaluation,and can only be use after confirming that all conditions are met.
If you prepare it yourself,you will easily damage the equipment because you don’t understand the material properties of the equipment,and the loss will outweigh the gain.Choosing a stable and high-quality special.
 If you prepare it yourself, you will easily damage the equipment because you don’t understand the material properties of the equipment, and the loss will outweigh the gain. Choosing a stable and high-quality special cleaning agent can not only maximize the potential of the equipment and prolong the service life of the equipment, but also ensure the stability and repeatability of the cleaning process to a greater extent.
This is because some components that rotate frequently, such as peristaltic pumps and their hoses, circulation pumps, etc., require regular inspection and replacement of accessories to ensure that the cleaning agent can be normally sucked according to the set value and the equipment can operate normally. Long-term shutdown may also cause some valves to fail or impurities to block pipelines. Such maintenance work can be done by internal equipment engineers, or outsourced to equipment manufacturers. Regular maintenance of the automatic cleaning machine is conducive to the long-term and efficient use of the equipment and the greater value of the equipment.
The specific maintenance also has the following points for attention, everyone must know:
1. Maintenance according to the use requirements of the bottle washing machine: For the sleeve roller chain, the bottle inlet system, the bottle outlet system, and the bearings of the return device, grease should be added once per shift; the drive shaft of the chain box, the universal coupling, etc. The other bearings are lubricated once every two shifts; the lubrication conditions of each gearbox are checked once a quarter, and the lubricating oil should be replaced when necessary.
2. Always pay attention to whether the movements of all parts are synchronized, whether there is any abnormal noise, whether the fasteners are loose, whether the liquid temperature and liquid level meet the requirements, whether the water pressure and steam pressure are normal, whether the nozzle and filter are blocked and cleaning, whether the bearing temperature is normal, and whether the lubrication is good. Once an abnormal situation is found, it should be dealt with in time.
3. Every time the washing liquid is changed and the waste water is discharged, the inside of the machine must be completely rinsed to remove dirt and broken glass, and the filter cartridge should be cleaned and dredged.
4. The heater should be sprayed with high-pressure water once a quarter, and the dirt filter and liquid level detector on the steam pipeline should be cleaned once.
5. Scrub the nozzles every month, dredge the nozzles, and adjust the alignment of the nozzles in time.
6. Check all kinds of chain tensioners every six months and adjust them when necessary.


Post time: Feb-06-2023